Some observations on the determination of germanium using hydride generation-Ultraviolet-Visible Molecular Absorption Spectrometry with a Diode- Array detector

  1. Sanz, J. 1
  2. Ortega, Luis A. 1
  3. Galban, J. 1
  4. Castillo, J.R. 1
  1. 1 Universidad de Zaragoza
    info

    Universidad de Zaragoza

    Zaragoza, España

    ROR https://ror.org/012a91z28

Revista:
Microchemical Journal

ISSN: 0026-265X

Año de publicación: 1990

Volumen: 41

Número: 1

Páginas: 29-40

Tipo: Artículo

Otras publicaciones en: Microchemical Journal

Resumen

In this paper a method for determining germanium by hydride generation-uv-visible molecular absorption spectrometry with diode-array detection is described. The molecular absorption maximum of germanium hydride appears to be outside the instrumental possibilities of the spectrophotometer and for this reason the determination is performed at 190 nm. Hydride generation and determination procedures are optimized and a linear response up to 200 μg is obtained. The method is used to determine germanium in an organogermanium compound. The sample is first halogenated and germanium is then extracted into NaOH aqueous solution. © 1990.